| Catalog Number |
100/597/080CZ |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3/h6: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that cuff was leaking.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
One used device was received for evaluation.Functional and visual tests were done, but the reported issue was duplicated.A leak was identified on the cuff.The root cause of the issue was unknown.A review of the device history record (dhr) indicated that all inspections were completed during manufacture, and no issues had been noted at that time.
|
| |
|