• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 101-9814
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
 
Event or Problem Description
Device analysis performed on the returned superion indirect decompression spacer revealed that the spindle cap was completely detached off from the implant body.The damage to implant indicated a failure, likely due to a combination of deployment against resistance (e.G., bone) and the physician failing to correctly attach the inserter to the spacer.Product analysis confirmed that the most probable cause of the complaint is unintended use error caused or contributed to event.Additionally, a labeling review was completed and revealed no anomalies.The instructions for use state that deployment should not be forced or implant breakage or damage to bony structures may result.Furthermore, if resistance to deployment is encountered, and repositioning of the implant is required, the driver should be rotated counterclockwise until a positive stop is reached and withdraw dorsally to collapse the cam lobes.
 
Event or Problem Description
Device analysis performed on the returned superion indirect decompression spacer revealed that the spindle cap was completely detached off from the implant body.The damage to implant indicated a failure, likely due to a combination of deployment against resistance (e.G., bone) and the physician failing to correctly attach the inserter to the spacer.Product analysis confirmed that the most probable cause of the complaint is unintended use error caused or contributed to event.Additionally, a labeling review was completed and revealed no anomalies.The instructions for use state that deployment should not be forced or implant breakage or damage to bony structures may result.Furthermore, if resistance to deployment is encountered, and repositioning of the implant is required, the driver should be rotated counterclockwise until a positive stop is reached and withdraw dorsally to collapse the cam lobes.
 
Additional Manufacturer Narrative
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Common Device Name
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
VERTIFLEX INC
2714 loker ave. west
carlsbad CA 92010
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key23430870
Report Number3006630150-2025-09814
Device Sequence Number12811416
Product Code NQO
UDI-Device Identifier00884662000550
UDI-Public(01)00884662000550(17)250417(10)800323
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P140004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 12/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2025
Device Model Number101-9814
Device Catalogue Number101-9814
Device Lot Number800323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2020
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer12/09/2025
Initial Report FDA Received Date10/30/2025
Supplement Report FDA Received Date12/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2485-2025
Patient Sequence Number1
Patient SexFemale
-
-