| Model Number |
V+C, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device needed service.Upon evaluation of the device, ventec observed that it was delivering low inspiratory pressure.There were no reports of patient use associated with the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of it delivering low inspiratory pressure was confirmed.Ventec replaced the blower to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined the cause of the reported issue to be the blower.
|
| |
|
Search Alerts/Recalls
|