PHILIPS ULTRASOUND, INC EPIQ DIAGNOSTIC ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
|
Back to Search Results |
|
| Model Number |
989605488741 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the transesophageal x8-2t transducer failed electrical leakage testing.The transducer was associated with the epiq cvx ultrasound system at the customer¿s site.The issue was discovered outside of clinical use.There was no patient or user harm associated with this event.The customer was provided with a replacement transducer to address their immediate concerns.Philips has started an investigation, and upon completion, a follow up report will be submitted.
|
| |
|