| Model Number |
101-9810 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
|
| |
|
Event or Problem Description
|
|
Device analysis performed on the returned superion indirect decompression spacer revealed the right wing of the superior cam lobe was slightly bent away from the median line.The implant did not function acceptably upon functional testing since the implant deployed with excessive resistance and one of the cam lobes would not retract without additional effort.The damage to implant indicates the observed damage was likely due to forced deployment against a rigid obstruction such as the spinous process.Additionally, a labeling review was completed and the instructions for use (ifu) states deployment should not be forced or implant breakage or damage to bony structures may result.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to the initial mdr in block b5.A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
|
| |
|
Event or Problem Description
|
|
Device analysis performed on the returned superion indirect decompression spacer revealed the right wing of the superior cam lobe was slightly bent away from the median line.The implant did not function acceptably upon functional testing since the implant deployed with excessive resistance and one of the cam lobes would not retract without additional effort.The damage to implant indicates the observed damage was likely due to forced deployment against a rigid obstruction such as the spinous process.Additionally, a labeling review was completed and the instructions for use (ifu) states deployment should not be forced or implant breakage or damage to bony structures may result.Furthermore, it states that if resistance to deployment is encountered, and repositioning of the implant is required, rotate the driver counterclockwise until a positive stop is reached and withdraw dorsally to collapse the cam lobes.
|
| |
|
Search Alerts/Recalls
|