Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation of this complaint has been completed.Our field service engineer investigated the system on-site and the reported failure was reproduced.The pressure transducer test failed.It was found that the reflector pressure transducer pcb was faulty and it was replaced.The reflector pressure transducer pcb was retained by the customer and not returned for investigation.The system device logs were received for evaluation.The evaluation of the received system logs confirms the reported system checkout failure.The system checkout failed the pressure transducer test due to the measured zero pressure offset for the reflector pressure transducer pcb being out of range; the measured zero pressure offset was 0 volt.The pressure transducer pcb has a factory calibrated zero pressure offset value that normally is 2 ± 0.5 v.During system checkout the zero-pressure offset value is measured and if the measured value is outside the allowed limit (±300 mv from factory calibrated value) the test will fail.The pressure transducer pcb is part of the pressure measuring in the system.A faulty pressure transducer pcb may lead to inaccuracy in pressure measurement.This will be detected during system checkout and high priority alarms will be generated if the failure occurs during treatment.Our conclusion is that the reported failure was caused by the replaced reflector pressure transducer pcb.
 
Event or Problem Description
It was reported that the anesthesia system failed in the pressure transducer test during system checkout.There was no patient involvement.Manufacturer's reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I C20
Common Device Name
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key23433718
Report Number8010042-2025-0001600
Device Sequence Number4782651
Product Code BSZ
UDI-Device Identifier07325710001349
UDI-Public(01)07325710001349(11)181004
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K191027
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6677200
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Initial Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-