| Model Number |
15 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Aborted Charge (2288)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.During initial evaluation of the customer¿s device it was observed that the device had logged an event code in the memory which is related to a potential issue of the device to deliver energy.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly dumped energy.During evaluation it was also observed that the device had logged an event code in the memory which is related to a potential issue of the device to deliver energy.In this state, the device may not be able to provide defibrillation therapy, if it were needed.This issue is patient related; however there was no adverse patient outcome reported.
|
| |
|