• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 36IN 4-0 D/A RB-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 36IN 4-0 D/A RB-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8557H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).Date sent to the fda: 10/31/2025 h6 component code: g07002 - pending evaluation of returned device this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: - did the reported issue contribute to any patient adverse consequences? - could you please provide the lot number? - please provide the source or name of person providing answers to follow-up questions: a device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that a patient underwent a pediatric heart surgery on an unknown date and suture was used.It was reported that dr.Said 7 of the sutures broke while trying to tie during a pediatric heart surgery.He didn¿t want to use any more from the box those 7 were pulled from.He was able to finish the procedure with suture from a different box.Device is available for return.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d4, h3, h4, h6.H3 investigation summary
=
> the product was returned to ethicon for evaluation.One open box with eleven (11) representative unopened samples was received for analysis.Product code 8557h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 36IN 4-0 D/A RB-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23436233
Report Number2210968-2025-12344
Device Sequence Number20970381
Product Code GAW
UDI-Device Identifier10705031017061
UDI-Public(01)10705031017061(17)291231(10)105KPH
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8557H
Device Lot Number105KPH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer11/03/2025
Initial Report FDA Received Date10/31/2025
Supplement Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-