|
SMITH & NEPHEW, INC. BIORAPTOR KNOTLESS SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72202403 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an unspecified procedure, the bioraptor knotless anchor was fractured during implantation.The procedure was resumed using an equivalent s+n back-up.It is unknown if there was a surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: part of the reported device was received for evaluation.A visual evaluation showed the device was returned with original packaging with the batch number of the complaint on the label.The anchor and plug were not returned.One green suture was returned.The insertion device has no visible defect.Bio debris is present.A material assessment could not be performed due to the part not being returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found one similar reported event.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include excessive force on the device, excessive torque on the device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|