| Model Number |
ENF-P4 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported, the rhino-laryngo fiberscope exhibited paint coming out of the scope.It was peeling, the paint particles where sticking to the customer's hand.The issue occurred during reprocessing.There were no reports of patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported, the rhino-laryngo fiberscope exhibited paint coming out of the scope.It was peeling, the paint particles where sticking to the customer's hand.The issue occurred during reprocessing.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Fields updated: h2, h3, h4, h6, h11 the device was returned to olympus for inspection.Based on the results of the investigation and historical data, reasonably known information suggests probable causes such as damage or deterioration of parts due to wear and tear, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|