|
ETHICON INC. SFXSYM PDS+ UNI VIO 24IN 1 SA OS-6; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
|
Back to Search Results |
|
| Catalog Number |
SXPP1A203 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a spine procedure on (b)(6) 2025 and barbed suture was used.The barbed suture broken during suturing in spine.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) date sent to the fda: 10/31/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: please confirm, how many devices broke during use on the patient? please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.Provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6.Investigation summary = > the product was returned to eth for evaluation.Visual inspection revealed that seven (7) unopened samples were received for analysis.Product code mcp4959h.Upon initial inspection, no external damages were observed on the packets.The packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand with no anomalies or issues related to breakage observed during evaluation.A functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|