Catalog Number 3503380
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 10/08/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with saline mentor smooth round high profile breast implants experienced deflation on the right side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent device removal.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On dec 5, 2025, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection, leak test and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant had a crease/fold on the posterior view.Leak testing was performed, according to mentor procedure, and it identified a tear within the crease/fold measuring approximately 0.2 cm.In addition, no more leak sites were observed.Microscopic examination was performed, and no striations were found.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND HIGH PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23438766
Report Number1645337-2025-12168
Device Sequence Number4782801
Product Code FWM
UDI-Device Identifier00081317001652
UDI-Public(01)00081317001652(17)260310(10)9702732
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2026
Device Catalogue Number3503380
Device Lot Number9702732
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/09/2025
Supplement Date Received by Manufacturer12/05/2025
Initial Report FDA Received Date10/31/2025
Supplement Report FDA Received Date12/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-