• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. BLUE RELOAD FOR ECHELON 60; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. BLUE RELOAD FOR ECHELON 60; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECR60B
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 10/31/2025.D4: batch #unk.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot 236c78, and no non-conformances were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.However, if the product is received at a later date, the investigation will be updated as applicable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a laparoscopic lobectomy procedure, it was observed that the staples were malformed on the third, fourth and fifth firing.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUE RELOAD FOR ECHELON 60
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key23438814
Report Number3005075853-2025-08477
Device Sequence Number4759085
Product Code GDW
UDI-Device Identifier10705036001775
UDI-Public(01)10705036001775(10)236C78(17)271231
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECR60B
Device Lot Number236C78
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2025
Initial Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-