| Model Number |
101-9808 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
|
| |
|
Event or Problem Description
|
|
Device analysis performed on the returned superion indirect decompression spacer revealed a severe abrasion on the mating surface of the actuator and severe scrapes were observed on one of the cam lobes.However, the implant functioned acceptably during the functional test.The damage to the implant indicates failure was likely due to deployment against resistance such as bone and/or manipulation of the position of the device by gear shifting of the inserter.Additionally, a labeling review was completed and reveal no anomalies.The instructions for use (ifu) states deployment should not be forced or implant breakage or damage to bony structures may result.It further warns against attempts to manipulate the position of the device with gross cranial/caudal/lateral articulation of the inserter, as the mechanical leverage provided by the length of the inserter may be sufficient to damage the device or surrounding anatomy.Application of excessive stress on instrumentation should be avoided.
|
| |
|
Event or Problem Description
|
|
Device analysis performed on the returned superion indirect decompression spacer revealed a severe abrasion on the mating surface of the actuator and severe scrapes were observed on one of the cam lobes.However, the implant functioned acceptably during the functional test.The damage to the implant indicates failure was likely due to deployment against resistance such as bone.Additionally, a labeling review was completed and revealed no anomalies.The instructions for use (ifu) states deployment should not be forced or implant breakage or damage to bony structures may result.Furthermore, it states that if resistance to deployment is encountered, and repositioning of the implant is required, rotate the driver counterclockwise until a positive stop is reached and withdraw dorsally to collapse the cam lobes.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Correction to the initial mdr in block b5.
|
| |
|