| Model Number |
CI-1601-05 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
10/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient's device was reportedly explanted due to pain and device migration.The recipient was reimplanted with another cochlear implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|