• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL SA UNIVERSAL ASEPTIC TRANSFER KIT HOUSING; BATTERY, REPLACEMENT, RECHARGEABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA UNIVERSAL ASEPTIC TRANSFER KIT HOUSING; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 89-8510-440-10
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2- foreign- spain.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the lid of aseptic transfer kit housing is broken.No consequences or impact to the patient.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h4, h6, h11.Complaint can be confirmed through the returned product analysis.Review of the most recent repair record determined broken hinge unit scrapped.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL ASEPTIC TRANSFER KIT HOUSING
Common Device Name
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key23448618
Report Number0008031000-2025-00188
Device Sequence Number12780955
Product Code MOQ
UDI-Device Identifier00889024355149
UDI-Public(01)00889024355149(11)230810(10)5019157
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number89-8510-440-10
Device Lot Number5019157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2025
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-