Correction: this mdr is being submitted to correct the submitted common device name under d2, model number, serial number, primary unique device identifier (udi) number, expiration date under d4 and manufacturing date under h4.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Complaint investigation results: a complaint investigation has been initiated to review the device history record (dhr) associated with the reported issue.The batch 6009171, in question was manufactured at the reynosa site.Device history record (dhr) review: the lot 6009171 was packaging according to work instruction (wi) version 116 in the line 07, on 09/sep/2024, with a total of (b)(4) units.Review of the dhr showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation was identified, nor maintenance events were recorded related to complaint code.Conclusion summary of complaint investigation: as a result of the following: no non-conformance (nc) raised during production related to complaint code, no trend identified for the lot in question and malfunction code, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.
|