PT. CIBA VISION BATAM ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during intraocular lens (iol) implant procedure, when the doctor attempted to load the iol (intraocular lens) into the cartridge after removing it from the packaging, the lens optic appears to have slight scratched.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., d.10., h3., h6., and in h11.The lens was returned anterior surface up in the lens case, positioned correctly.Dried viscoelastic was observed on both sides of the lens.The lens was cleaned with klrp for further evaluation.Scratches were observed at the center of both sides of the lens.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Associated products were not provided.It is unknown if qualified associated products were used.The company lens model is only qualified for use in the company cartridges with the company handpiece.The reported lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on observation, and the review of manufacturing records, it is concluded that the damage is not manufacturing related.Due to the presence of blood, surgical solution, and the condition of the returned sample, the damage is most likely related to customer handling.It is unknown if qualified associated products were used.The company lens model is only qualified for use in the company cartridges with the company handpiece.The instructions for use (ifu) instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in h3, h6, and h11 the lens was returned anterior surface up in the lens case, positioned correctly.Dried viscoelastic was observed on both sides of the lens.The lens was cleaned with klrp for further evaluation.Scratches were observed at the center of both sides of the lens.A batch record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the batch record were met.The account indicated the use of a qualified associated cartridge with a non-qualified viscoelastic.The handpiece used was not provided.It is unknown if a qualified handpiece was used.The reported lens damage was observed.The root cause is most likely related to a failure to follow the instructions for use (ifu).The account indicated the use of a qualified associated cartridge with a non-qualified viscoelastic.The handpiece used was not provided.It is unknown if a qualified handpiece was used.The company lens model is only qualified for use in the company cartridges with the company handpiece.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
PT. CIBA VISION BATAM
beringin kav. #204 batamindo
industrial park muka kuning
pulau batam 29433
ID  29433
Manufacturer (Section G)
PT. CIBA VISION BATAM
beringin kav. #204 batamindo
industrial park muka kuning
pulau batam 29433
ID   29433
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23449578
Report Number9681121-2025-00150
Device Sequence Number18156874
Product Code HQL
UDI-Device Identifier00380652552608
UDI-Public(01)00380652552608(17)290116(11)240119(10)31567929
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.240
Device Lot Number31567929
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer01/04/2026
03/06/2026
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date01/30/2026
04/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AUROGEL PLUS.; MONARCH II IOL, CARTRIDGE B.
Patient SexUnknown
-
-