| Model Number |
MA60AC |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during intraocular lens (iol) implant procedure, when the doctor attempted to load the iol (intraocular lens) into the cartridge after removing it from the packaging, the lens optic appears to have slight scratched.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., d.10., h3., h6., and in h11.The lens was returned anterior surface up in the lens case, positioned correctly.Dried viscoelastic was observed on both sides of the lens.The lens was cleaned with klrp for further evaluation.Scratches were observed at the center of both sides of the lens.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Associated products were not provided.It is unknown if qualified associated products were used.The company lens model is only qualified for use in the company cartridges with the company handpiece.The reported lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on observation, and the review of manufacturing records, it is concluded that the damage is not manufacturing related.Due to the presence of blood, surgical solution, and the condition of the returned sample, the damage is most likely related to customer handling.It is unknown if qualified associated products were used.The company lens model is only qualified for use in the company cartridges with the company handpiece.The instructions for use (ifu) instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h3, h6, and h11 the lens was returned anterior surface up in the lens case, positioned correctly.Dried viscoelastic was observed on both sides of the lens.The lens was cleaned with klrp for further evaluation.Scratches were observed at the center of both sides of the lens.A batch record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the batch record were met.The account indicated the use of a qualified associated cartridge with a non-qualified viscoelastic.The handpiece used was not provided.It is unknown if a qualified handpiece was used.The reported lens damage was observed.The root cause is most likely related to a failure to follow the instructions for use (ifu).The account indicated the use of a qualified associated cartridge with a non-qualified viscoelastic.The handpiece used was not provided.It is unknown if a qualified handpiece was used.The company lens model is only qualified for use in the company cartridges with the company handpiece.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
|
| |
|