| Model Number |
DS460TBT |
| Medical Device Problem Codes |
Degraded (1153); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted as part of a batch submission of complaints deemed reportable following record correction and remediation efforts.
|
| |
|
Event or Problem Description
|
|
A device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of foam degradation.During the evaluation, burnt power connector was observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service centre.There was 3 error code found.The third-party service centre concludes that they couldn't confirm the customer's allegation and there were no visible foam degradation and unit got corrected.
|
| |
|
Search Alerts/Recalls
|