• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED; SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED; SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 123614C
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the patient presented with extensive cerebral infarction and urinary incontinence.On (b)(6) 2025, an indwelling urinary catheter was placed per medical orders to address urinary incontinence.Throughout (b)(6) 2025, catheter drainage remained unobstructed with no leakage.At 8:51 am on (b)(6) 2025, during urethral opening care, leakage from the urinary foley catheter was discovered, soaking the bedding.Immediate care was initiated.Examination revealed a ruptured catheter balloon causing leakage.The attending physician and head nurse were immediately notified.At 10:20 am on (b)(6) 2025, a new 14fr urinary catheter was inserted per medical orders, and urine drainage was unobstructed.
 
Event or Problem Description
It was reported that the patient presented with extensive cerebral infarction and urinary incontinence.On (b)(6) 2025, an indwelling urinary catheter was placed per medical orders to address urinary incontinence.Throughout (b)(6) 2025, catheter drainage remained unobstructed with no leakage.At 8:51 am on (b)(6) 2025, during urethral opening care, leakage from the urinary foley catheter was discovered, soaking the bedding.Immediate care was initiated.Examination revealed a ruptured catheter balloon causing leakage.The attending physician and head nurse were immediately notified.At 10:20 am on (b)(6) 2025, a new 14fr urinary catheter was inserted per medical orders, and urine drainage was unobstructed.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure could be for this failure mode could be user related (example: contact with sharp object/exposure to petrolatum based products mechanical failure/operator error).The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action required at this time.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Common Device Name
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23453127
Report Number1018233-2025-09805
Device Sequence Number12780989
Product Code KOD
UDI-Device Identifier00801741095184
UDI-Public(01)00801741095184(11)240729(17)290328(10)MYJT5102
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number123614C
Device Lot NumberMYJT5102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2025
Supplement Date Received by Manufacturer11/05/2025
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-