MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® URINE METER WITH BAG; DRAIN BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® URINE METER WITH BAG; DRAIN BAG Back to Search Results
Catalog Number 153204
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that immediately after opening the product the sterilization pack, discovered that the connection between the inlet tube and the meter of the drain bag was broken.
 
Event or Problem Description
It was reported that immediately after opening the product the sterilization pack, discovered that the connection between the inlet tube and the meter of the drain bag was broken.
 
Additional Manufacturer Narrative
The reported event was confirmed ¿ cause unknown.The product had caused the reported failure.Sample was evaluated and observed connecting part between urine meter and inlet tube was broken that allowed water to leak out.The exact cause of how and when the problem occurred could not be determined however, there was an existing scar (b)(4) issued and documented to supplier for further investigation.A potential root cause for this failure could be defective / torn / broken components and also might be contributed by inappropriate material selection, inadequate solvent selection or improper mating dimensions.The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Corrections: d, f, g, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URINE METER WITH BAG
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23453316
Report Number1018233-2025-09830
Device Sequence Number8056674
Product Code KNX
UDI-Device Identifier00801741033117
UDI-Public(01)00801741033117(11)250515(17)270628(10)MYKN7042
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number153204
Device Lot NumberMYKN7042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Supplement Date Received by Manufacturer02/04/2026
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date02/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-