SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 5MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72204390 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a midfoot ligament insertion, after the insertion site was prepared with a 5 mm femoral drill bit, when placing the screws using an orange handle, even after tapping with a 6 mm tap, a 5x25 mm biosure screw fractured.There were no fragmented pieces left inside the patient.The procedure was resumed after a non-significant surgical delay with a change in surgical technique by fixing the tendon using super suture.A void was left in the patient.No further complications were reported, and the patient is stable.
|
| |
|
Search Alerts/Recalls
|
|
|