| Model Number |
72077-01 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
Loss of consciousness (2418)
|
| Date of Event |
10/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An incompatible application issue was reported with the abbott diabetes care (adc) device in use with iphone 12 with ios operating system and version 18.6.2.As a result, customer was unable to monitor glucose readings due to the device showing signal loss and no glucose alarms.The customer experienced loss of consciousness in the middle of the street and hit their face with something and passersby called ambulance for them.The customer received unspecified third-party treatment.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is pending at this time.A follow up will be submitted once additional information is obtained.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An incompatible application issue was reported with the abbott diabetes care (adc) device in use with iphone 12 with ios operating system version 18.6.2 with software version 3.6.5.12633.As a result, customer was unable to monitor glucose readings due to the device showing signal loss and no glucose alarms.The customer experienced loss of consciousness in the middle of the street and hit their face with something and passersby called ambulance for them.The customer received unspecified third-party treatment.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The user reported signal loss and no alarms due to incompatible application issue.The reported issue was investigated.The reported configuration was not compatible with the customer's reported application.The compatibility guide was available to the customer on the product's website.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Search Alerts/Recalls
|