| Catalog Number |
21-2766-24 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that cannot inject the solution.There is a blockage.This occurred during priming in early (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
One used device was received for evaluation.Functional and visual tests were done, the reported issue was duplicated.A solvent membrane was observed in the inner diameter of the needle.The root cause of the issue was an errant solvent application error.A review of the device history record (dhr) indicated that all inspections were completed during manufacture, and no issues had been noted at that time.
|
| |
|