MAUDE Adverse Event Report: COOK INC NCOMPASS NITINOL TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NCOMPASS NITINOL TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G46027
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the ncompass nitinol tipless stone extractor¿s basket snapped in the duct during a laparoscopic trans-cystic common bile duct exploration procedure.The basket tip was retrieved during the same procedure.Although the procedure could not be completed, and a single stone was left behind.An endoscopic retrograde cholangiopancreatography (ercp) was needed to remove the remaining stone and the patient had a prolonged hospital stay.A section of the device did not remain inside the patient¿s body.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1 - phone number: (b)(6).G4 ¿ pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation: it was reported the ncompass nitinol tipless stone extractor¿s basket snapped in the duct during a laparoscopic trans-cystic common bile duct exploration procedure.The basket tip was retrieved during the same procedure.Although the procedure could not be completed, and a single stone was left behind.An endoscopic retrograde cholangiopancreatography (ercp) was needed to remove the remaining stone and the patient had a prolonged hospital stay.A section of the device did not remain inside the patient¿s body.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.The complaint device was not returned; therefore, no functional testing or visual inspections could be performed.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for lot.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field.Cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: - 'intended use: the ncircle, ncompass, and nforce stone extractors are intended for stone manipulation and removal in the urinary tract.' based upon the available information, no product returned, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCOMPASS NITINOL TIPLESS STONE EXTRACTOR
Common Device Name
FFL, DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key23458017
Report Number1820334-2025-01358
Device Sequence Number15424898
Product Code FFL
UDI-Device Identifier00827002460270
UDI-Public(01)00827002460270(17)280617(10)16647319
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG46027
Device Catalogue NumberNCT4-017115
Device Lot Number16647319
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2025
Supplement Date Received by Manufacturer01/08/2026
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date01/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1