| Model Number |
M00545180 |
| Medical Device Problem Codes |
Break (1069); Deformation Due to Compressive Stress (2889); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) biliary duct cannulation procedure to treat jaundice and bile duct stricture due to lithiasis performed on (b)(6) 2025.During the procedure, the cutting wire of the autotome rx 44 completely broke at the proximal end after stretching.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed still completed with the same original device.There were no patient complications reported as a result of this event.Note: photo of the complaint device provided by the customer outside the patient shows the cutting wire broken and kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Imdrf device code a0406 captures the reportable event of cutting wire kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Imdrf device code a0406 captures the reportable event of cutting wire kinked.Block h11: (investigation result).The returned autotome rx 44 was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Media analysis also observed the same problems.No other problems with the device were noted.The reported event of cutting wire break and wire kinked were confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) biliary duct cannulation procedure to treat jaundice and bile duct stricture due to lithiasis performed on (b)(6) 2025.During the procedure, the cutting wire of the autotome rx 44 completely broke at the proximal end after stretching.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed still completed with the same original device.There were no patient complications reported as a result of this event.Note: photo of the complaint device provided by the customer outside the patient shows the cutting wire broken and kinked.
|
| |
|