SMITH & NEPHEW, INC. AHTB UNIVERSAL HIP DISTRACTOR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Catalog Number |
72203270 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a hip arthroscopy, during the early portion of the procedure, the traction of the universal hip distractor was lost.Attempts to re-traction the joint failed.Upon further inspection, it was noted the carriage knob of the operative side mast was broken.Specifically, a pin sheared that attached the knob to the bolt that tightens the distractor boot assembly to the mast.The procedure was completed with a surgical greater than 30 minutes using an equivalent sn back-up device.There was prolonged time under anesthesia.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|