• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN TA; STAPLER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN TA; STAPLER, SURGICAL Back to Search Results
Model Number TA6048S
Medical Device Problem Codes Failure to Fire (2610); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
Date of Event 10/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a procedure involving the open stapler, the tissue was not cut and staple line bleeding occurred.The surgeon had to resect and re-staple with a manual stapler.
 
Event or Problem Description
It was reported that during left colectomy, involving the open stapler , the tissue was not cut and staple line bleeding occurred.The surgeon had to resect and re-staple with a manual stapler.
 
Additional Manufacturer Narrative
Additional information: g3 correction: b5.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23460448
Report Number2647580-2025-03270
Device Sequence Number18717111
Product Code GAG
UDI-Device Identifier10884521043237
UDI-Public(01)10884521043237(17)290930(10)P4K1316(20)01
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K231491
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTA6048S
Device Catalogue NumberTA6048S
Device Lot NumberP4K1316
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Supplement Date Received by Manufacturer11/03/2025
Initial Report FDA Received Date11/03/2025
Supplement Report FDA Received Date11/03/2025
Date Device Manufactured10/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-