| Model Number |
0282 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a high out of range gradual rising pacing impedances.It was also noted that there were gradual rising shock impedances within normal limits and noisy signals.Technical services (ts) recommended to have a lead revision.Tachy therapy was stated to be turned off for this patient.This lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated b5.Describe event or problem with additional information obtained from the field.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a high out of range gradual rising pacing impedances.It was also noted that there were gradual rising shock impedances within normal limits and noisy signals.Technical services (ts) recommended to have a lead revision.Tachy therapy was stated to be turned off for this patient.This lead remains in service.No adverse patient effects were reported.Additional information was received from the field.This lead was explanted and replaced.This product has not been returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of "cause traced to device design".Updated b5.Describe event or problem with additional information obtained from the field.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a high out of range gradual rising pacing impedances.It was also noted that there were gradual rising shock impedances within normal limits and noisy signals.Technical services (ts) recommended to have a lead revision.Tachy therapy was stated to be turned off for this patient.This lead remains in service.No adverse patient effects were reported.Additional information was received from the field.This lead was explanted and replaced.This product has not been returned for analysis.No additional adverse patient effects were reported.
|
| |
|