| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
|
| Date of Event |
10/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a knee arthroplasty revision of the articular surface to address infection accompanied by pain and swelling approximately one (1) month post-operatively.Attempts have been made, however, no additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - concomitant devices - persona cemented stemmed tibial component right size d catalog #: 42532006702 lot #: 66974233, persona cruciate retaining standard femoral component right size 7 catalog #: 42502606202 lot #: 67070944, persona cemented all-poly patella 32mm catalog #: 42540200032 lot #: 67142152.G2 - report source - foreign: event occurred in australia.The complainant has not yet indicated whether the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, g3, g6, h2, h6, h11.The following sections were corrected: h4.The reported event could not be confirmed as medical records were not provided.The device history records were not reviewed as the event was determined to be unrelated to the implanted device.Sterile certifications were reviewed and found to be conforming with no applicable deviations.The device was verified to have gone through acceptable sterilization processes following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors that may include hospital/surgical environment, provider-related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issue affecting implant safety or effectiveness, the implanted products are not identified as the source or contributing to the reported infection.The root cause was determined to be traced to non-device related factors.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|