Philips has investigated this complaint.According to additional information collected, the system was not in clinical use when the issue was identified.It was reported to philips that fluro and cine are not functioning.Review of the logfile shows generator reports showing a sign of a low-quality hospital mains power.Upon troubleshooting, the philips remote service engineer (rse) checked the system remotely and found that ¿generator busy¿ messages appeared intermittently.The philips field service engineer (fse) checked the system onsite and found that the system was functioning normally, and checked the mains power, which was found to be satisfactory.The fse kept the system under observation for few days and rechecked the system, including mains voltage and earthing, and confirmed that it was operating properly.No service has been performed by philips.To date, no further information was received.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, it was identified that the issue occurred without patient involvement and was identifiable prior to procedure commencement, which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use, and the issue happened due to improper maintenance.Based on re-evaluation, this case is not reportable.The codes were updated based on the investigation outcome.
|