This mdr is being submitted as part of a retrospective review of records dating january 2022-december 5, 2024, under capa 10308384.The late submission of this report is justified by the thorough and detailed nature of the retrospective review process.This process was essential to accurately capture all relevant data and ensure the integrity of our reporting.We have included the capa reference for the retrospective review in the additional manufacturer narrative section of the fda form 3500a, as recommended.A review of the complaint history for sn (b)(6) was performed in salesforce which did locate 1 similar complaint with the same failure mode for this serial number.¿ case: (b)(4).¿ evening coverage needed pyxis medstation 4000 orders are not deleting.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 11-apr-2012 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device patient orders failed to be crossed from the interface and the customer stated that due to this issue they were unable to receive medication for patients.A technical support specialist restarted carefusion coordination engine services and confirmed that the messages were crossed.The system functioned as intended after the technical support specialist troubleshot the device.
|