• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number E7507
Medical Device Problem Codes Difficult to Remove (1528); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Burn(s) (1757); Peeling (1999); Insufficient Information (4580)
Date of Event 10/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: e7507-db, e7507-db polyhesive ii rem ret el w/cord (lot#unknown); e7507-db, e7507-db polyhesive ii rem ret el w/cord (lot#unknown); e7507, e7507 polyhesive ii rem ret el w/cord (lot#251500024t).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that after the procedure, the patient experienced skin peeling that was considered a burn after using the pad.There was a black circular spot found in one of the pads as shown in the pictures provided.
 
Additional Manufacturer Narrative
Additional information: information received shows that this event is a duplicate of another issue reported under regulatory report#:1717344-2025-01425.This file will be closed and all further updates processed under the above-referenced report for this event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
D10 concomitant product: e7507, e7507 polyhesive ii rem ret el w/cord (lot # 251810287t); e7507-db, e7507-db polyhesive ii rem ret el w/cord (lot # unknown); e7507, e7507 polyhesive ii rem ret el w/cord (lot # 251500024t).Additional information: d3 (street 1, mfr city and mfr region), d4 (model number, catalog number, expiration date, lot number and udi), g3, h4.Correction: h6 (rfr, fdp coding).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN
boulevard insurgentes 19030
tijuana,mx CO 80301 3299
Manufacturer (Section G)
COVIDIEN
boulevard insurgentes 19030
tijuana,mx CO 80301 3299
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23464787
Report Number1717344-2025-01429
Device Sequence Number8861406
Product Code GEI
UDI-Device Identifier10884524000077
UDI-Public(01)10884524000077(17)270430(10)251500024T(20)01
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K822572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE7507
Device Catalogue NumberE7507
Device Lot Number251500024T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2025
Supplement Date Received by Manufacturer11/03/2025
11/12/2025
Initial Report FDA Received Date11/04/2025
Supplement Report FDA Received Date11/11/2025
11/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H11
Patient SexUnknown
-
-