| Model Number |
PESSARY |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Codes |
Abrasion (1689); Genital Bleeding (4507)
|
| Date of Event |
10/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is in process.A supplemental follow-up report will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
Consumer reported that the applicator petals of an unspecified number of pessaries were partially open and during insertion she was pinched and noticed blood and a cut.She has improved.She did not seek medical attention and she did not report any serious adverse health effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Samples received from consumer were consistent with the reported complaint.The investigation was completed.The root cause of the reported event was determined to be related to process controls during manufacturing at the petal forming station.At this time, the affected product has been discontinued, and no further manufacturing will be undertaken.If production is resumed in the future, improved process controls will be implemented at the petal forming station prior to manufacturing.Additional information: g3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: if follow-up, what type? h6: investigation findings, investigation conclusion.H11: manufacturer narrative.
|
| |
|