| Catalog Number |
168-300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation" and "exchange from textured to smooth due to the patients concerns with the product".This record is for the right side.The device was explanted, replaced and will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported events anxiety - product/ procedure and deflation was received on december 09,2025 with lot number 1336802.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification (if applicable).Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ deflation: observed an opening through microscopic inspection assessed as [assessment].No further actions are required as it is not related to the manufacturing process.¿ anxiety - product/ procedure : unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (crease and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation" and "exchange from textured to smooth due to the patients concerns with the product".This record is for the right side.The device was explanted, replaced and will be returned.
|
| |
|