| Model Number |
768901 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that before use, the device was experiencing o2 leakage failure and the locking tab on power knob is non-functional.Complainant indicated that there was no patient involvement in the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer at the customer site detected and verified a leak originating from the pressure switch in the pneumatics compartment by taking off the cover and sensing the airflow.Technical support advised the switch pressure / electrical panel mounting required replacement to resolve the reported issue.A replacement order was placed.
|
| |
|