| Model Number |
ILS-1800-KT |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, biopsy forceps were inserted into the extended working channel and, upon removal, abraded plastic strips from the inner extended working channel layer were found balled in the forceps jaw along with tissue.The extended working channel was then removed, and the forceps were reinserted, but no tissue was gained.The procedure was completed without navigation.The abraded plastic strips were removed using forceps and suction with irrigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Correction: b1, h5, h8.H3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted string like fibers.The string like fiber was determined to be from scraping the inner liner.Functional testing noted that the lg and ewc lengths had been measured using calibrated rulers and were found to be within specification.The ewc had been checked with a ball mandrel and was observed to pass without any resistance.The ewc had been cut open to inspect the interior, where scrapes/gouges were observed.It was reported that there was string like plastic found from the inner extended working channel.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: with the jaws closed, insert the distal tip of the device into the extended working channel.Advance the device using short, 2cm strokes.When the biopsy operation is complete, ensure that the jaws remain closed and slowly withdraw the forceps from the extended working channel.If the jaws will not close completely, do not use force to pull the jaws into the extended working channel.Remove the endoscope and the extended working channel with the forceps still in place.Then manually close the jaws and remove the forceps and extended working channel from the endoscope.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|