(b)(4).This report is being submitted to relay additional information.The following fields have been updated: b4, b5, d4 (udi#), g3, g6, h2, h3, h4, h6, h11.The reported event has been confirmed.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.Review of the provided photos/returned product found the root cause of the reported event can be attributed to transit damage.The root cause of the reported event can be attributed to transit damage and a packaging design issue.A corrective action has been previously raised to implement improved packaging changes.These packaging changes will reduce the number of transit related packaging damage events.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|