• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH BIOLOX® DELTA, CERAMIC FEMORAL HEAD, M, ø 32/0, TAPER 12/14; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH BIOLOX® DELTA, CERAMIC FEMORAL HEAD, M, ø 32/0, TAPER 12/14; PROSTHESIS, HIP Back to Search Results
Catalog Number 00877503202
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 10/09/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10.Durasulâ®, alpha insert, hooded, ii/32 item# 0100013509 lot# 3166836.Fitmoreâ®, shell with fins, uncemented, 52/ii item# 0100024452 lot# 3028067.Unknown stem item# unknown lot# unknown.G2.Report source: australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
It was reported that the patient had an initial hip replacement, subsequently, approximately 1 year and 8 months post implantation, underwent a revision due to dislocation.Diligence is completed and no further information on the reported event has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.No product was returned or pictures provided; a product evaluation could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Complaints are monitored per complaint trending process in order to identify potential adverse trends.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
Diligence is completed and no further information on the reported event has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX® DELTA, CERAMIC FEMORAL HEAD, M, ø 32/0, TAPER 12/14
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23468565
Report Number0009613350-2025-00854
Device Sequence Number8791134
Product Code LZO
UDI-Device Identifier00889024430327
UDI-Public(01)00889024430327(17)321116(10)3141133
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00877503202
Device Lot Number3141133
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date11/04/2025
Supplement Report FDA Received Date01/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexUnknown
-
-