• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DURATA STS OPTIM PASSIVE FIXATION; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DURATA STS OPTIM PASSIVE FIXATION; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 7170/65
Medical Device Problem Codes Fracture (1260); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Additional information was received that technical support was contacted and there were increased pacing impedance episodes observed on the rv lead.
 
Event or Problem Description
During an in-clinic follow-up, noise resulting in oversensing was observed on the right ventricular (rv) lead.Lead conductor fracture was suspected but it was not confirmed.A revision procedure was performed.The rv lead was capped and replaced to resolve the event.The patient is stable with no consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM PASSIVE FIXATION
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23468666
Report Number2017865-2025-1004078
Device Sequence Number8791151
Product Code LWS
UDI-Device Identifier05414734502436
UDI-Public(01)05414734502436(10)4130999(17)160630
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2016
Device Model Number7170/65
Device Lot Number4130999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Supplement Date Received by Manufacturer11/06/2025
Initial Report FDA Received Date11/04/2025
Supplement Report FDA Received Date11/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/21/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CURRENT PLUS VR.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-