| Model Number |
7170/65 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information was received that technical support was contacted and there were increased pacing impedance episodes observed on the rv lead.
|
| |
|
Event or Problem Description
|
|
During an in-clinic follow-up, noise resulting in oversensing was observed on the right ventricular (rv) lead.Lead conductor fracture was suspected but it was not confirmed.A revision procedure was performed.The rv lead was capped and replaced to resolve the event.The patient is stable with no consequences.
|
| |
|