As reported by an edwards lifesciences affiliate, approximately 7 years and 3 months post implantation of a 23mm sapien xt valve in the pulmonic position, a valve in valve was performed with a 23mm sapien 3 ultra resilia valve due to moderate pulmonic valve regurgitation, mean gradient is 20mmhg w/mild ps.Patient symptoms: worsening rv dilation via mri.
|
|
A supplemental mdr is being submitted due to engineering evaluation findings.Sections: g3, g6, h2, h6 type of investigation, investigation findings, investigation conclusions have been updated.The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.Additional assessment of the failure mode is not required at this time.The device was not returned for evaluation.Due to the unavailability of the complaint device, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.The instructions for use (ifu), and training manuals have been reviewed.No inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported regurgitation was confirmed based on available information to date.The reported event does not allege a malfunction that could be related to an edwards manufacturing deficiency and/or one was not confirmed through investigation.A review of the ifu and training manuals revealed no deficiencies.During the manufacturing process, all sapien xt valves are 100% visually inspected for defects and 100% tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.Per the instructions for use (ifu), valve regurgitation, including central regurgitation and paravalvular leak are potential adverse event associated with bioprosthetic heart valves and the tavr procedure.Regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration, including calcification, non-calcific degeneration, leaflet thickening or fibrosis, or a combination of these.Regurgitation may also develop progressively if host fibrotic tissue, or pannus, grows onto the bioprosthetic valve.Pannus, a cause of nonstructural dysfunction, may interfere with functionality of the device by restricting the leaflet motion leading to abnormal coaptation.In addition, paravalvular leak refers to blood flowing through a channel between the structure of the implanted valve and the cardiac tissue, as a result of a lack of appropriate sealing of the valve to the target site.Some pvl is not uncommon post deployment.Many cases are mild to moderate, and either resolve over time or do not cause symptoms.Others may be more clinically significant and require intervention.The mechanism behind worsening or late pvl is not well understood but may be related to cardiac remodeling.Due to limited information provided, a definitive root cause was unable to be determined at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.Since no edwards product defects or labeling deficiencies were identified, no corrective or preventative action is required.
|