| Model Number |
DTMB1QQ |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that a remote monitor reader position error occurred with the cardiac resynchronization therapy defibrillator (crt-d), resulting in the monitor not connecting and no telemetry with mycarelink.Troubleshooting steps included adjusting the position of the reader for placement errors, power cycling the monitor, and removing any interference within six feet.The patient was referred to the clinic.The device remains in use. no patient complications have been reported as a result of this event.
|
| |
|