U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE FISHER AND PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE FISHER AND PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850AGU
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in germany reported that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Section g4: the mr850agu is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for the similar product is k110019.The subject mr850 respiratory humidifier has been requested to be returned to fisher & paykel healthcare new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Corrections: section b4:date of this report updated.Section g4: the mr850agu is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for the similar product is k110019.Method: the speaker from the subject mr850 respiratory humidifier was received at fisher & paykel (f&p) healthcare in new zealand for investigation, where it was performance tested.Our investigation is thus based on the evaluation of the subject device, the information provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: performance testing of the pcb from the subject mr850 respiratory humidifier confirmed that no sound was generated from the speaker.Resistance testing identified an open circuit in the speaker's coil.Conclusion: the root cause of the speaker failure was due to an open circuit in the speaker coil.The mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
A healthcare facility in germany reported via a fisher & paykel healthcare (f&p) field representative, that the audible alarm of an mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER AND PAYKEL HEALTHCARE
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23472493
Report Number9611451-2025-01314
Device Sequence Number10228521
Product Code BTT
UDI-Device Identifier09420012407238
UDI-Public(01)09420012407238(10)2102146397(11)220422
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR850AGU
Device Catalogue NumberMR850AGU
Device Lot Number2102146397
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2025
Supplement Date Received by Manufacturer02/19/2026
Initial Report FDA Received Date11/04/2025
Supplement Report FDA Received Date03/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-