MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 54MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 54MM Back to Search Results
Catalog Number 940-02-54G
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Health Effect - Clinical Codes Unspecified Infection (1930); Septic loosening (4884)
Date of Event 10/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that due to implant loosening and infection patient required revision surgery.
 
Additional Manufacturer Narrative
Once the investigation is completed, any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
See d4 expiration date, h4 h6, h10, and h11.The agent reported "(implant loosening, infection)".The previous surgery and the surgery detailed in this event occurred 2 years and 10 months apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for review.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to an infection and loosening.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 54MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key23472678
Report Number1644408-2025-01650
Device Sequence Number10252549
Product Code LPH
UDI-Device Identifier00190446311577
UDI-Public00190446311577
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number940-02-54G
Device Lot Number801Z1217
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer12/17/2025
Initial Report FDA Received Date11/04/2025
Supplement Report FDA Received Date01/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
402-03-282 LOT CODE 85881151; 951-01-44G LOT CODE 2218A1006; 952-28-44G LOT CODE 2226A1001
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-