• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS CORPORATION GLIDELIGHT LASER SHEATH; DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS CORPORATION GLIDELIGHT LASER SHEATH; DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS Back to Search Results
Model Number 500-302
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 10/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B6/b7) patient information regarding relevant tests/laboratory data or medical history - not available from facility.H3/h6) the device was not returned, thus no investigation could be completed, and the cause of the reported failure could not be confirmed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
A lead extraction procedure commenced to remove a right ventricular (rv) and a right atrial (ra) lead due to cied system/pocket infection.Spectranetics lld lead locking devices (llds) were inserted into each lead to provide traction.While lasing through calcified lesions, the spectranetics 14f glidelight laser sheath outer jacket separated, and fibers were exposed.A new glidelight was used to complete the procedure with no reported patient harm.This event is being reported for unintended radiation exposure, potential for harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDELIGHT LASER SHEATH
Common Device Name
DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS
Manufacturer (Section D)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
nicole worford
9965 federal drive
colorado springs, CO 80921
719447-2465
MDR Report Key23472874
Report Number3007284006-2025-00232
Device Sequence Number21464368
Product Code MFA
UDI-Device Identifier00813132020330
UDI-Public(01)00813132020330(17)260318(10)FGB24C08A
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
P960042
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2026
Device Model Number500-302
Device Catalogue Number500-302
Device Lot NumberFGB24C08A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2025
Initial Report FDA Received Date11/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PHILIPS LASER SYSTEM; RA PACING LEAD (MANUFACTURER, MODEL UNK); RV LEAD (MANUFACTURER, MODEL, TYPE UNK); SPECTRANETICS LLD LEAD LOCKING DEVICES
Patient SexMale
-
-