MAUDE Adverse Event Report: SOPHYSA PRESSIO ICP MONITORING KIT, PARENCHYMAL WITH BOLT; ICP MONITORING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOPHYSA PRESSIO ICP MONITORING KIT, PARENCHYMAL WITH BOLT; ICP MONITORING CATHETER Back to Search Results
Model Number PSO-PB
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
Icp catheter replacement following accidental sectioning.After insertion of the new catheter, it was connected to the monitor.The monitor began to reboot.The monitor and cables were replaced several times, without success.Icp catheter replacement was successful but required a new cranial incision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSIO ICP MONITORING KIT, PARENCHYMAL WITH BOLT
Common Device Name
ICP MONITORING CATHETER
Manufacturer (Section D)
SOPHYSA
5 rue guy môquet
orsay, 91400
FR  91400
Manufacturer (Section G)
SOPHYSA
rue sophie germain
technopôle temis
besançon, 25000
FR   25000
Manufacturer Contact
julie lopez
5 rue guy môquet
orsay, 91400
FR   91400
MDR Report Key23476514
Report Number3001587388-2025-00112
Device Sequence Number18724429
Product Code GWM
UDI-Device Identifier03760124130768
UDI-Public(01)03760124130768(11)250411(17)300201
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K162108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Pharmacist
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 06/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberPSO-PB
Device Catalogue NumberPSO-PB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2025
Supplement Date Received by Manufacturer10/22/2025
Initial Report FDA Received Date11/05/2025
Supplement Report FDA Received Date06/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
CAPA 2025-04
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
Patient SexMale
-
-