Zimvie did not receive one (1) unknown driver for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr, sterilization, and complaint history review could not be performed, as the subject lot number associated with the unknown driver is not available.Zimvie quality management system (qms) has controls in place to prevent the distribution of non-conforming product and ensure the product is within specifications.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is incorrect techniques used.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up" h6: entered evaluation codes.H10: added manufacturer narrative.
|