• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 ALYTE¿ Y-MESH GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 ALYTE¿ Y-MESH GRAFT Back to Search Results
Catalog Number Y500
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that 3mm mesh extrusion on posterior wall vagina.Mri pelvis requested and starting vaginal oestrogen.
 
Additional Manufacturer Narrative
The reported event could not be confirmed.No sample was returned for evaluation.Based on the investigation findings, current manufacturing controls are considered adequate as to detect and segregate any non-conforming unit.Event described could not be confirmed as a manufacturing related issue.This occurrence does not introduce a new hazard or harm.It represents a known failure mode that has been assessed in the product risk management file in accordance with applicable regulatory standards and internal procedures.The associated risks continue to be monitored as part of ongoing post-market activities in compliance with both regulatory requirements and local procedures.Product labeling was reviewed for this investigation.The reported event is addressed within the labeling as it states: warnings: the effectiveness of this product has not been validated by a prospective randomized clinical trial.The implant procedure carries an inherent risk of infection and bleeding, as do similar urological procedures.After use, the product and its packaging should be treated as a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state, and federal laws and regulations.Precautions: based on physician experience and training, a thorough assessment of each patient should be made to determine their suitability for a synthetic mesh procedure.Additional consideration should be given to the use of alyte y-mesh graft in patients with a compromised immune system, any condition that would compromise healing, or any patient with a history of prior abdominal or pelvic surgeries.Consideration should also be given to the ability of the patient to tolerate the surgical procedure.Accepted surgical practice and precautions must be followed for the management of infected or contaminated wounds.Postoperative bleeding may occur in some patients and must be controlled prior to patient release.Sutures should not be placed in the mesh edge.Sutures should be placed a minimum of 1cm from any mesh edge.Inadequate suturing of the graft material to the pelvic tissues may lead to failure of the repair, additional complications and recurrence of prolapse.Check the integrity of the packaging before use.Do not use the mesh if the packaging is opened or damaged.As for any implantable material, it is recommended to open the package at the time of implantation.The alyte y-mesh graft is intended as a single use device.Do not re-sterilize any portion of the alyte y-mesh graft.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Post-operatively the patient should be advised to refrain from heavy lifting, exercise (e.G.Cycling, jogging) and/or intercourse until the physician determines it is suitable for the patient to return to normal activities.Adverse events: complications associated with the proper implantation of the alyte y-mesh graft may include, but are not limited to those typically associated with surgically implantable materials including: postoperative hematoma, seroma, abscess or fistula formation, or scarring which may occur following the implant procedure.Urinary retention, bladder outlet obstruction and other voiding and defecatory dysfunctions.These conditions may be associated with over-correction/too much tension placed on the implant.Perforations or lacerations of vessels, nerves, bladder, bowel, urethra, rectum, or any viscera, which may occur during the implantation procedure.Irritation at the operative wound site which may elicit a foreign body response that leads to wound dehiscence, inflammation and/or infection.Extrusion through vaginal epithelium or erosion into surrounding viscera and/or mucosa.Inflammation, sensitization, pain, dyspareunia, scarification, contraction, device migration and failure of the procedure resulting in recurrence of vaginal wall prolapse.Urinary incontinence (stress and urge).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: d.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that 3mm mesh extrusion on posterior wall vagina.Mri pelvis requested and starting vaginal oestrogen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALYTE¿ Y-MESH GRAFT
Common Device Name
Y-MESH
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23476928
Report Number1018233-2025-09917
Device Sequence Number13310927
Product Code OTO
UDI-Device Identifier00801741016165
UDI-Public(01)00801741016165(11)131026(17)151028(10)HUXJ1162
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K101722
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2015
Device Catalogue NumberY500
Device Lot NumberHUXJ1162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer02/10/2026
Initial Report FDA Received Date11/05/2025
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/26/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-