| Catalog Number |
STS-OM-001 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an applicator deployment occurred.The sensor was inserted into the abdomen on (b)(6) 2025.The sensor pod was provided for investigation.An external visual inspection was performed and failed due to missing sensor wire.Applicator deployment was not found within the investigation window.The allegation was not confirmed.However, detached or missing sensor wire was found in connection to the reported allegation.The probable cause of the detached or missing sensor wire could not be determined.No injury or medical intervention was reported.
|
| |
|