RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number ES2100X15B
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to a third-party service center for preventive maintenance.There was no harm or injury reported.During the evaluation of the device at a third-party service center, an oxygen proportional valve stuck alarm was observed in the device's downloaded event log.The device's oxygen blending module was replaced to address the issue.Additionally, a low inlet pressure alarm was observed.This would not affect the device's ability to deliver therapy as designed.The oxygen blending module replacement would address this issue also.The device's air inlet foam filter and particulate filter were replaced due to preventive maintenance protocol.The device was cleaned and tested and passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23479382
Report Number2518422-2025-111605
Device Sequence Number10499497
Product Code CBK
UDI-Device Identifier00606959054097
UDI-Public(01)00606959054097
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberES2100X15B
Device Catalogue NumberES2100X15B
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2025
Initial Report FDA Received Date11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-